You are seeing this message because your Web browser does not support basic Web standards. Find out more about why this message is appearing and what you can do to make your experience on this site better.


ABOUT ARCHIVES
Advanced Search

Welcome   | My Account | E-mail Alerts | Access Rights | Sign In


  Vol. 134 No. 10, October 1998 TABLE OF CONTENTS
  Archives
  •  Online Features
  Study
 This Article
 •Full text
 •PDF
 •Send to a friend
 • Save in My Folder
 •Save to citation manager
 •Permissions
 Citing Articles
 •Citing articles on HighWire
 •Citing articles on ISI (9)
 •Contact me when this article is cited
 Related Content
 •Similar articles in this journal
 Topic Collections
 •Surgery
 •Laser Surgery
 •Dermatologic Laser Surgery
 •Dermatologic Procedures, Other
 •Facial Plastic Surgery
 •Cosmetic Surgery/ Procedures
 •Alert me on articles by topic

Optimizing Fluence and Debridement Effects on Cutaneous Resurfacing Carbon Dioxide Laser Surgery

Noah Kawika Weisberg, MD; Timothy Kuo; Berhooz Torkian; Lou Reinisch, PhD; Darrel L. Ellis, MD

Arch Dermatol. 1998;134:1223-1228.

Objective  To develop methods to compare carbon dioxide (CO2) resurfacing lasers, fluence, and debridement effects on tissue shrinkage and histological thermal denaturation.

Design  In vitro human or in vivo porcine skin samples received up to 5 passes with scanner or short-pulsed CO2 resurfacing lasers. Fluences ranging from 2.19 to 17.58 J/cm2 (scanner) and 1.11 to 5.56 J/cm2 (short pulsed) were used to determine each laser's threshold energy for clinical effect. Variable amounts of débridement were also studied.

Main Outcome Measures  Tissue shrinkage was evaluated by using digital photography to measure linear distance change of the treated tissue. Tissue histological studies were evaluated using quantitative computer image analysis.

Results  Fluence-independent in vitro tissue shrinkage was seen with the scanned and short-pulsed lasers above threshold fluence levels of 5.9 and 2.5 J/cm2, respectively. Histologically, fluence-independent thermal depths of damage of 77 µm (scanner) and 25 µm (pulsed) were observed. Aggressive debridement of the tissue increased the shrinkage per pass of the laser, and decreased the fluence required for the threshold effect. In vivo experiments confirmed the in vitro results, although the in vivo threshold fluence level was slightly higher and the shrinkage obtained was slightly lower per pass.

Conclusions  Our methods allow comparison of different resurfacing lasers' acute effects. We found equivalent laser tissue effects using lower fluences than those currently accepted clinically. This suggests that the morbidity associated with CO2 laser resurfacing may be minimized by lowering levels of tissue input energy and controlling for tissue debridement.


From the Department of Otolaryngology (Dr Reinisch), Vanderbilt University Medical Center (Dr Weisberg and Messrs Kuo and Torkian), Nashville, Tenn; and the Department of Medicine, Division of Dermatology, Vanderbilt University and Nashville Veterans Affairs Medical Centers, Nashville (Dr Ellis).



THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES

Exogenous Administration of Substance P Enhances Wound Healing in a Novel Skin-Injury Model
Delgado et al.
Exp. Biol. Med. 2005;230:271-280.
ABSTRACT | FULL TEXT  

Why Does Carbon Dioxide Resurfacing Work?: A Review
Ross et al.
Arch Dermatol 1999;135:444-454.
ABSTRACT | FULL TEXT  